Description / Abstract:
This part of ISO 7206 describes a method of determining the
torque required, under specified laboratory conditions, to loosen
the fixation of the head of hip joint prostheses in which the head
is not intended to rotate relative to the neck. It applies to the
femoral component of total or partial hip joint replacements in
which the head and neck/stem are secured together by a locking
conical taper or any other means and in which the head and
neck/stem are separate components, and which are made of metallic
and non-metallic materials.
This part of ISO 7206 does not cover methods of examining the
test specimens; these should be agreed between the test laboratory
and the party submitting the specimen for test.